Detector non-linearity corrections are applied to Classic Cam, IRCAM,
and INGRID. For IRCAM the correction is
times the
square of the bias-subtracted signal. For INGRID the
measured-to-actual counts is given by the expression
1.0 - 1.2247
x 10-6 * M - 7.68045 x 10-11 * M2, where M is
the measured ADU count, although it is actually applied after the
pre-exposure readout is subtracted from the post-exposure integration.
For Classic Cam the correction is 1.625
* (1 +
overhead/exposure_time), where the overhead is 2 * speed *
number of pre-exposure reads + (speed + readout time ) * number
of post-exposure reads. The various parameters in this expression are
either read from the headers or taken verbatim from an IRAF script cited
in the user manual.
[CLASSICCAM/_CORRECT_NONLINEARITY_,
IRCAM/_CORRECT_NONLINEARITY_,
INGRID/_CORRECT_NONLINEARITY_PRE_POST_]